7. Preparing moisture analyser to work
During heating, the halogen heaters 1 warm up to very high
temperature. When drying chamber is opened avoid touching the
heaters as it may cause severe burns or damage the heaters!
Dryer chamber cover 3 heats up to 40°C, but perforated cover may
heat up over 60°C. Do not touch the top cover during drying as it may
cause severe burns!
1
2
3
7
6
5
4
1 – halogen heaters
2 – temperature sensor
3 – chamber cover
4 – dryer chamber
5 – pan plate
6 – single-use pan grip
7 – single-use pa
8 – distance stake
9 – regulated legs
10 – level
8
10
9
1. Take all contents out of a package: the moisture analyser and packed separately: the
tin pan shield, single use pans, the pan handle and the pan support. It is
recommended to keep the original moisture analyzer package in order to transport
the moisture analyzer safely in future
2. Place the moisture analyzer on a stable ground not affected by mechanical vibrations
and airflows.
3. Level the moisture analyzer with rotating legs 9 so that the air bubble in water-level
10 at the back of the moisture analyzer is in the middle and the moisture analyser
rests on all four legs.
4. Open the dryer chamber with the handle at the front. Put the pan shield 4 on three
distance sleeves 8. Gently insert the pan support 5 into the mechanism hole.
5. Place a single use pan 7 on the pan handle 6 and put the pan on the pan support
(the handle should rest on the pan shield so that it does not touch the pan or the pan
support).
6. Close the drying chamber cover 3 and plug the device to the mains (230V).