8
USER MANUAL
7. After self-tests and result stabilisation zero indication is displayed. The dryer starts
initial heating (signalised with an appropriate communicate). After initial heating the
moisture analyser is ready to work.
When temperature during initial heating exceeds 105°C or heating
time is longer than 1 minute, terminate initial heating with CLR key and
check if the temperature sensor 2 works properly and if both halogen
heaters light 1 (see chapter 15).
In case any defect occurs contact an authorised service point.
8. The moisture analyzer should not be used to weigh ferromagnetic materials due to
accuracy decrease.