10-2
Chapter 10.  MAINTENANCE AND TROUBLESHOOTING
Alarm displays and corrective action
The following table shows the alarm displays and corrective actions if any failure 
occurs in this unit:
Alarm code
Failure name Cause Corrective action
PV input failure 
(Over-range)
Sensor burnout, incorrect 
wiring, incorrect PV input 
type setting
Check the wiring. 
Set the PV input type again.
PV input failure 
(Under-range)
Sensor burnout, incorrect 
wiring, incorrect PV input 
type setting
CJ failure Terminal temperature is 
faulty (thermocouple).
Check the ambient 
temperature.
PV input failure Sensor burnout, incorrect 
wiring (RTD)
Check the wiring.
RSP input failure 
(Over-range) 
(Displayed only in 
the RSP mode.)
Sensor burnout, incorrect 
wiring, incorrect RSP input 
type setting
Check the wiring.  
Set the RSP input type again.
RSP input failure 
(Under-range) 
(Displayed only in 
the RSP mode.)
Sensor burnout, incorrect 
wiring, incorrect RSP input 
type setting
Check the wiring. 
Set the RSP input type again.
MFB input failure Burnout, incorrect wiring Check the wiring. 
Check the MFB input value. 
Check the settings for 
C6
 
1
 
and 
62
 in the setup bank.
Motor adjustment 
failure
Check for burnout or 
incorrect wiring. 
Motor power shutdown.
Readjust the motor after 
checking the wiring and 
motor power. 
Check the settings for 
C6
 
1
 
and 
62
 in the setup bank. 
p.5-109 (for details on 
AL
 
10
)
CT input failure 
(over-range)  
(CT input 1 or 2, or 
both)
A current exceeding the 
upper limit of the display 
range was measured. The 
number of CT turns or the 
number of CT power wire 
loops is incorrectly set, or 
wiring is incorrect.
Use a CT with the correct 
number of turns for the 
display range, reset the 
number of CT turns, reset 
the number of CT power 
wire loops, and/or check the 
wiring.
A/D conversion 
failure
A/D converter is faulty. Replace the unit.
Parameter failure Data is corrupted by noise, 
or power is shut-down while 
the data is being set.
• Restart the unit.
• Set the data again (set 
data for 
AL95
/
97
 and 
adjustment data for 
AL96
/
98
).
• Replace the unit.
Adjustment data 
failure
Data is corrupted by noise, 
or power is shut-down while 
the data is being set.
Parameter failure 
(RAM area)
Data is corrupted by noise.
Adjustment data 
failure (RAM area)
Data is corrupted by noise.
ROM failure ROM (memory) is faulty. • Restart the unit.
• Replace the unit.
  Handling Precautions
•  If ROM version 1 of the instrument information bank (
I
d02
) is prior to 
2.04, CT input failure (
AL
 
1
 
1
) is not displayed.
•  If 
AL07
 and 
AL
 
10
 are displayed alternately, take corrective actions for 
AL
10
 first.
•  This device will not recover from 
AL07
 and 
AL
 
10
 by restarting it. 
Corrective actions are described in the above table.