EasyManua.ls Logo

Baker Panametrics PanaFlow XMT1000 - Namur Ne107

Default Icon
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PanaFlow™ XMT1000 User’s Manual 109
Appendix F. Foundation Fieldbus Communication
To Edit the fields in the Transducer Blocks, complete the following steps:
1. Select or enter the new value.
2. Click on the Write Changes button.
3. Return to the Resource Block > Others tab and select Commit from the SYTEM_REQ_COM drop down box
4. Click on the Write Changes button.
To Exit configuration mode, complete the following steps:
1. Select Cancel from the SYSTEM_REQ_CMD drop down box.
2. Click on the Write Changes button.
Note: The XMT1000 will automatically exit configuration mode after 5 minutes of inactivity.
F.4.3 NAMUR NE107
The NAMUR NE107 recommendation specifies that detailed device-specific diagnostics are summarized as four
simple status signals. The diagnostics are set to defaults by Panametrics, but they can be modified to any other level
by the user. The four status signals are:
Failed: This category is typically used for hardware or software failures. The meter output is not valid. Consult
with the factory for a resolution.
Offspec: This category is typically used for application, installation, or process problems. Consult the
troubleshooting section of this appendix or contact Panametrics customer service for assistance.
Check: This category means the output of the device is invalid due to on-going work on the device, such as
programming, etc.
Maintenance: This category is typically used to assign parameters that are in good status but which may fall
out of specification due to some process condition or wear factor. There are no diagnostics set as default in
this category.
The status signals (see Figure 71 below) can either be reported as errors (ACTIVE) or masked when the error occurs
(MASK). There are no default settings for the MASK bits.
Figure 71: NAMUR NE107 Configuration in Resource Block

Table of Contents