PanaFlow™ XMT1000 User’s Manual 125
Appendix F. Foundation Fieldbus Communication
F.12 Simulation Mode
Simulation mode allows the user to test the FF implementation without the instrument providing real data. The meter
PCB is shipped with simulation mode disabled. To enable simulation mode, complete the following steps:
1. Remove the PCB from the meter.
2. Locate the jumper P5 (see Figure 79 below).
3. Move the P5 jumper to the left one place (pins 2 & 3) to enable simulation mode.
4. Re-install the PCB into meter
5. Verify that the Block error field in the Resource Block shows SimulationActive status.
Figure 79: P5 on XMT1000 PCB
CAUTION!
To prevent damage to the electronic components, always use ESD protection whenever
handling printed circuit boards.
Jumper P5