PanaFlow™ XMT1000 User’s Manual 33
Chapter 3. Programming
Figure 30: Modbus Port Setting
3.5.2 Standard Analog Output
The XMT1000 has one Analog Output and one Digital Output in Standard configuration.
3.5.2.1 Setting up Analog Output
The XMT1000 meter has one Analog Output in standard configuration. For additional Analog outputs Optional I/O
boards may be purchased. Use steps as in section “Log-in and Primary Pages” to navigate to the Input/Output
settings page.
1. Then highlight [Analog Output (Std)] and press [ENTER].
2. If you do not wish to connect an Analog Output, you should turn Analog Output selection to OFF.
3. If you are connecting an Analog Output, choose 4-20mA option. The Figure 31 below, shows the options available.
4. Select the Measurement to be sent out on the 4-20mA output, followed by the [Base Value] and [Full Value]
selection. Refer Table 4 to see measurement options available for Analog output.
5. Select [Error Handling]. Refer to section 3.5.2.2 to choose an option that suits your needs.
Table 4: Measurement Options for Analog output
Measurement Channel Measurement Options for Analog Output
Composite Velocity, Actual Volumetric, Standard Volumetric, Mass Flow, Average
Volumetric Flow Rate, Sound speed, Reynolds#
Channel x Velocity, Actual Volumetric, Standard Volumetric, Mass Flow, Average
Volumetric Flow rate, Sound speed, Standard Deviation of Gain, Gain and
SNR