PanaFlow™ XMT1000 User’s Manual 13
Chapter 2. Installation
Figure 12: XMT1000 Terminal Board Wiring Diagram
2.7.1 Wiring the Analog Outputs
The standard configuration of the XMT1000 flow transmitter includes one isolated 4-20 mA analog output.
Connections to this output may be made with standard twisted-pair wiring, but the current loop impedance for this
circuit must not exceed 600 ohms. Two additional analog outputs are available as an option.
To wire the analog outputs, complete the following steps:
1. Disconnect the main power to the unit and remove the wiring cover.
2. Install the required cable gland in the chosen conduit hole on the side of the electronics enclosure.
3. Refer to Figure 12 on page 13 for the location of the terminal block and wire the analog output as shown. Secure
the cable gland.