Calibration procedure (continued)
6.3.2 Adjusting
The detailed procedure for adjusting is
described on page17 ff.
The factory set null point and end point is replaced by a
new start point and end point and the associated output
values can be adjusted. The start and end values can be
adjusted as you like up to the limits.
Adjustment is possible from serial number
120615000xxxxx xx onwards.
Steps
► Move magnet to the new start position.
► Read new start point by pressing the buttons.
► Set the new start value by pressing the buttons.
New null point
Before
After
Fig. 6-4: Adjust a new start value
► Move magnet to the new end position.
► Read new end point by pressing the buttons.
► Set the new end value by pressing the buttons.
New end point
Before
After
Fig. 6-5: Adjust a new end value
6.3.3 Reset
Restoring the transducer to its factory settings.
The detailed procedure for the reset is
described on page19.
New start value
New end value