B&C Electronics CL 7685
Operator’s Manual - Rev. C - 05/05
- 21
[MODE] to go to
----------------------------------
_
D5 |AL 0.0/20.0ppm| Alarm parameters display
0.0 PPM: lower value limit
20.0 PPM: higher value limit
[CAL] to activate the alarm values programming sequences
[MODE] to go to
----------------------------------
_
D6 |CLEANING OFF | Cleaning function display
CLEANING OFF: cleaning function disabled
MANUAL CLEAN: manual cleaning function
AUTO CLEAN: automatic cleaning function
[CAL] to activate the cleaning function programming sequences of relay D
[MODE] to go to
----------------------------------
_
D7 |01 10.0mA/1.0 pp| Analog output N°1 / ppm value
[MODE] to go to
----------------------------------
_
D7BIS |01 10.0mA/1.0 pp| (Option 091.3711)
Analog output N°2 / (PPM value or temp. value)
[MODE] to go to
----------------------------------
_
D8 |Configuration | Configuration display
[CAL] - to activate the keyboard lock/unlock and LCD display contrast selection
sequences
- to activate the configuration sequences
[MODE] to go to
----------------------------------
_
D9 |CL7685 R:2.3x
| P/N and Software release
[MODE] to go back to the main Display D0