14
Site~Rite
*
5
Ultrasound System
Precautions (continued)
Caution: Attach power source in such a way as to prevent damage. Improper installation may
damage power cords.
Caution: Inspect the probe prior to each use. If damage to the cable or transducer face is noted, do
not use the probe. Damage to the system may occur.
Caution: Hot water (in excess of 113
o
F or 45
o
C) may damage the probe.
Caution: Use only Bard Access Systems cleaning and disinfection procedures. Failure to do so
may damage the device.
Notes
Note: When cleaning the system and components, it is important to remove all particles or other
matter from all surfaces and crevices.
Note:
For 240 V applications use only center tapped 240 VAC single phase power.