31
Instructions For Use
9 Technical Specifications
9.1 Operating and Storage Conditions
Operating Temperature: 59°F to 100°F (15°C to 38°C)
Storage Temperature: 50°F to 95°F (10°C to 35°C)
Operating Humidity: 5% to 85% Relative Humidity (non-condensing)
Storage Humidity (packaged): 5% to 95% Relative Humidity (non-condensing)
Storage Humidity (unpackaged): 5% to 85% Relative Humidity (non-condensing)
9.2 Scanner Specifications
Dimensions: 12" W x 12" H x 5" D
Weight: 10 lbs.
Power Sources: AC adapter, Internal and External DC Battery Pack
Power Consumption: 82 Watts Maximum
Monitor Size:
12.1” diagonal
IEC 60601- 1: Class II, Type BF Applied Part, Continuous Operation, Internally Powered
Equipment, Not Category AP or APG Equipment, Not protected against
ingress of water.
9.3 Probe Acoustic Output Specifications
Description
of Probe
Operating
Mode
I
spta.
X
(X denotes statisti-
cally determined
maximum)
FDA I-
spta
.3
Published
Values
MI X
(X denotes
statistically
determined
maximum)
FDA MI
Published Values
L-VA B 78.2mW/cm
2
Peripheral Vessel <
720 m W/cm
2
Cardiac <
430 mW/cm
2
Fetal Imaging & Other** <
94 mW/cm
2
1.1
Peripheral Vessel <
1.9
Cardiac <
1.9
Fetal Imaging & Other** <
1.9