User’s Guide92
8. Install the second sample tube onto the SIP.
9. Set the current tube pointer to Beads_002.
10. Click Acquire Data to begin acquisition.
11. Before recording, preview the data on the MyData worksheet
to verify that all expected populations are visible and the data
is similar to the previous sample.
12. Click Record Data.
13. When event recording has completed, remove the second tube
from the cytometer.
14. If you are recording more than two tubes, repeat steps 8
through 13 for the remaining tubes.
15. Print the experiment-level cytometer settings by right-clicking
the Cytometer Settings icon in the Browser and selecting Print.
16. Install a tube of DI water onto the SIP.
17. Place the cytometer in standby mode.
More information Analyzing data (page 93)