FBXXX Field Temperature Block
Serial Commands - Alphabetic Listing
50
This command is a read only command and returns the state of the reference function-
ality (0, 1).
CAL:DATE:MOD[?] [<year>,<month>,<day>] (-R model only)
NOTE: This command is unconditionally protected, which requires a password
to set it.
Read or set the calibration date for the -R Module (INPUT CALibration DATE) where
the entered values are all numeric and “yyyy” is a four digit year (2000-2135), “mm”
is a two digit month (1-12), and “dd” is a two digit day (1-31)
Read Example: CAL:DATE:MOD?
Response: 2007,05,24
Set Example: CAL:DATE:MOD 2007,12, 30
This command reads or sets the calibration date for the -R module including all ac-
companying sensors.
CAL:DATE:UNIT[?] [<year>,<month>,<day>]
NOTE: This command is unconditionally protected, which requires a password
to set it.
Read or set the calibration date for the Main Unit where the entered values are all nu-
meric and “yyyy” is a four digit year (2000-2135); default 2007, “mm” is a two digit
month (1-12), and “dd” is a two digit day (1-31)
Read Example: CAL:DAT:UNIT?
Response: 2007,05,24
Set Example: CAL:DAT:CAL 2006,12, 30
This command reads or sets the calibration date for the Main Unit. Calibration is for
the heat source portion of the instrument.
CALC1:CONV:CAT? (-R model only)
Read the list of reference probe characterization methods, “CVD, I90, IEC, RES”
Example: CALC1:CONV:CAT?
Response: “CVD”,”I90”,”IEC”,”RES”
Provides the list of PRT/RTD characterization methods available.