51
Digital communication interface
Serial Commands - Alphabetic Listing
CALC1:CONV:DATE[?] [<yyyy,mm,dd>] (-R model only)
NOTE: This command is conditionally protected, which requires a password to
set it.
Read or set the Smart Lemo probe calibration date in yyyy,mm,dd format
Year Range = {2000 – 2135}; Default: 2007
Month Range = {1 – 12}; Default: 1
Day Range = {1 – 31}; Default: 1
Read Example: CALC1:CONV:DATE?
Response: 2007,10,09
Set Example: CALC1:CONV:DATE 2007,09,06
This command reads or sets the calibration date for the Smart Lemo probe.
CALC1:CONV:NAME[?][n] (-R model only)
NOTE: This command is conditionally protected, which requires a password to
set it.
Read or set the reference probe characterization method where “n” is an alpha
numeric value; CVD, I90, IEC, and RES. Range={ITS90, CVD, IEC, RES} or 0-3,
Default:ITS90
Read Example: CALC1:CONV:NAME?
Response: CVD
Set Example: CALC1:CONV:NAME I90
Reads or sets the reference probe characterization method from a predened set of
options.
CALC1:CONV:PAR:CAT? (-R model only)
Read the list of active reference probe characterization parameter names
Example: CALC1:CONV:PAR:CAT?
Response: “RTPW”,”A”,“B”,“C”,“A4”,“B4”
This command is a read only command, which returns the active parameters for the
current reference probe type.