FBXXX Field Temperature Block
Serial Commands - Alphabetic Listing
52
CALC1:CONV:PAR:VAL[?] par[,<n>] (-R model only)
NOTE: This command is conditionally protected, which requires a password to
set it.
Read or set a reference probe characterization parameter. Where “par” is a parameter
identied as follows: ITS90: RTPW, A7, B7, C7, A4, B4; or CVD: AL, DE or BE. “n”
may be some real number or an exponential value such as -1.234567e-5 for ITS-90
ITS-90 Ranges: {RTPW, A7, B7, A4, B4} or 0 - 5
RTPW = 1.0 to 200 ohms
Coefcients = ± 0.010
Defaults: RTPW = 100
All ITS-90 coefcients = 0.00000
CVD Ranges/Defaults: {R0, AL, DE, BE} or 0 - 3
R0 = 1.0 to 200.00 ohms
R0 Default = 100.00
AL Range = 0.1 to 0.9
AL Default = 0.00385055
BE Range = 0.0 to 1.0
BE Default = 0.10863
DE Range = 0.0 to 2.0
DE Default = 1.499786
Read Example: CALC1:CONV:PAR:VAL? RTPW
Response: 100.4545
Set Example: CALC:CONV:PAR:VAL A7, 0.00385075
This command returns a reference probe characterization parameter as desired by the
user.
CALC1:CONV:PROG (-R model only)
NOTE: This command is unconditionally protected, which requires a password
to set it.
Program the Smart Lemo probe with the current probe settings
Example: CALC1:CONV:PROG