Digital communication interface
Serial Commands - Alphabetic Listing
CALC1:CONV:SNUM[?] <ser_num> (-R model only)
NOTE: This command is conditionally protected, which requires a password to
set it.
Read or set the reference probe serial number, where “ser_num” character range
equals {0-9, A-Z, ‘-‘}, up to 10 characters in length with a minimum of 1 character
length. Default: “0”
When a blank space is entered, any characters after the blank are dropped. For ex-
ample, change S/N 1234-5678 to S/N TEST1. Enter TEST1<Blank Space>678. The
serial number will drop the last three characters and enter the S/N TEST1.
Read Example: CALC1:CONV:SNUM?
Response: 1234
Set Example: CALC1:CONV:SNUM 1560-D
This command allows the user to read or enter a reference probe serial number.
CALC1:CONV:TEST[?] <n>]
Test the conversion algorithm from resistance (ohms) to temperature (°C or °F). “n” is
the value to be converted (ohms); entry for “n” is required for converted output. Range
= {0-500}; Default = 100
Read Example: CALC1:CONV:TEST? 100.000
Response: 0.0100
This is a read only command and tests conversion algorithm from resistance to
temperature.
CAL1:DATA? (-R model only)
Reads the reference sensor Temperature. The returned value will be in degrees °C
(°F) if a temperature value is returned. The value may be resistance depending on the
conversion selection
Example: CALC1:DAT?
Response: 325
The command returns an instantaneous reading of the reference sensor Temperature.
OUTP:STAT[?] [0|1]
Read or set the Main Heat output enable, off [0] or on [1]
Read Example: OUTP:STAT?
Response: 0
Set Example: OUTP:STAT 1