FBXXX Field Temperature Block
Serial Commands - Alphabetic Listing
54
This command reads or sets the active heating or cooling output status. A “0” is re-
turned if the output status is off, and a “1” is returned if the output status is on
OUTP1:DATA?
Read the main heat output percent
Example: OUTP1:DATA?
Response: 18.0
This command returns the current main zone heater duty cycle
OUTP2:DATA?
Read the gradient heat output percent
Example: OUTP2:DATA?
Response: 57.0
This command returns the current top zone heater percent power.
PROG:CAT?
A catalog list of all dene programs: Ramp & Soak = SEQ
Example: PROG:CAT?
Response: “SEQ”
PROG:IDEN[?] [n]
Read or set the program identier. Character range = {0 – 9, A – Z, ‘-‘}, up to 12 char-
acters, minimum 1 character: Default: “0”
Read Example: PROG:IDEN?
Set Example: TEST-1
PROG:MEM:CLE (-R Model only)
NOTE: This command is unconditionally protected, which requires a password
to set it.
Erase all test reports stored in NVMemory
Example: PROG:MEM:CLE
PROG:MEM:COUN? (-R Model only)
Read the test report count.
Example: PROG:MEM:COUN? 6