FBXXX Field Temperature Block
Serial Commands - Alphabetic Listing
56
If the selected program is not running then a value of 0 is returned otherwise a 1 is
returned.
PROG:TYPE[?] [ <prog>]
Read or select a program to run where “prog” is a name, SEQ. Default = SEQ
Read Example: PROG:TYPE?
Response: SEQ
Set Example: PROG:TYPE SEQ
Reads or selects the current program setting, Ramp & Soak = SEQ.
READ?, MEAS? or FETC? (-R model only)
Read the Reference sensor temperature, °C or °F
Example: READ?
Response: 264.262
If the external reference probe is enabled, the reference temperature is returned other-
wise 0.0 is returned.
SENS1:CAL:PAR<n>[?][cal] (-R model only)
NOTE: This command is unconditionally protected, which requires a password
to set it.
Read or set a reference input calibration parameter where “n” is a value of 1 or 2 cor-
responding to the calibration parameters REF1C0 and REF1C100 respectively. “cal” is
a real number used as the calibration offset for the respective parameter
REF1C0 Range = {-1.0 to 1.0}
REF1C100 Range = {-2.0 to 2.0}
Defaults (all): 0.0000
Read Example: SENS:CAL:PAR1?
Response: 0.2
Set Example: SENS1:CAL:PAR2 0.092
Reference thermometer input commands to verify or set REF1C0 (PAR1) or RE-
F1C100 (PAR2) calibration parameters.
SENS1:DATA? (-R model only)
Read the reference input resistance
Example: SENS1:DATA?