57
Digital communication interface
Serial Commands - Alphabetic Listing
Response: 199.9366
This command returns the resistance in ohms of the reference probe.
SOUR:LCON:DER[?] [n]
NOTE: This command is unconditionally protected, which requires a password
to set it.
Read or set the main control loop derivative time in seconds, Min: 0.0, Max: 99.9
Read Example: SOUR:LCON:DER?
Response: 1.5
Set Example: SOUR:LCON:DER 5
The main zone derivative is the derivative time in seconds that the instrument’s PID
controller uses for main zone control.
SOUR:LCON:INT[?] [n]
NOTE: This command is unconditionally protected, which requires a password
to set it.
Read or set the main control loop integral time in seconds. Range = {10.0-999.9}
Read Example: SOUR:LCON:INT?
Response: 20.0
Set Example: SOUR:LCON:INT 10
The main zone integral is the integration time in seconds that the instrument’s PID
controller uses for main zone control.
SOUR:LCON:PBAN[?] [n]
NOTE: This command is unconditionally protected and requires a password to
set it.
Read or set the main control loop proportional band, °C. Range = {1.0-99.9}
Read Example: SOUR:LCON:PBAN?
Response: 1.5
Set Example: SOUR:LCON:PBAN 7
The main zone proportional band is the gain in °C that the instrument’s proportional-
integral-derivative (PID) controller uses for main zone control.