FBXXX Field Temperature Block
Serial Commands - Alphabetic Listing
58
SOUR:LIST:SPO<i>[?] [n]
Read or set a main temperature preset set-point
Read example: SOUR:LIST:SPO6?
Response: 25.00
Set Example: SOUR:LIST;SPO6 100.00
This sets the preset set-points found in PROG MENU under RAMP/SOAK.
SOUR:PROT:HCUT?
Read the hard cutout temperature set-point in °C or °F
Read Example: SOUR:PROT:HCUT?
Response: 140
Returns the current value of the hard cutout set-point.
SOUR:PROT:CLEA
Reset the cutout to enable the system
Example: SOUR:PROT:CLEA
This command has no response.
If the Field Temperature Block exceeds the temperature set in the soft cutout menu or
if it exceeds the maximum operating temperature of the instrument, a cutout condition
occurs. If this happens, the instrument enters cutout mode and will not actively heat
or cool until the user issues this command to clear the cutout or resets the instrument
using the Setpt. key to clear the cutout mode and activate the instrument.
SOUR:PROT:SCUT:LEV[?] [n]
NOTE: This command is conditionally protected and requires a password to set
it.
Read or set the soft cutout set-point where “n” is an integer value from 0 to 700
FB150 Range = {–25.00 to 165.00}
FB350 Range = {25.00 to 365.00}
FB660 Range = {25.00 to 670.00}
Read Example: SOUR:PROT:SCUT:LEV?
Response: 125
Set Example: SOUR:PROT:SCUT:LEV 450
Read or set the soft cutout set-point. The soft cutout should be set to protect the tem-
perature limits of the instruments under test.