Manual for VMP300-based instruments
97
Maximum Current
resolution
0.004 % of the range (down to 76 aA on the ± 1 pA)
Applied Current
Accuracy
< ±0.1% of range ±0.03% of setting for ±500 mA to ±100 nA ranges
< ±0.1% of range ±1% of setting for ±10 nA range to ±1 nA ranges
< ±0.2% of range ±2% of setting for ±100 pA range
Current Measurement
Ranges
± 100 pA, ± 1 nA, ± 10 nA, ± 100 nA
Additional Ranges
with gain
± 1 pA, ± 10 pA
Maximum Resolution
0.004 % of the range
Accuracy
< ±0.1% of range ±0.03% of setting for ±500 mA to ±100 nA ranges
< ±0.1% of range ±1% of setting for ±10 nA range to ±1 nA ranges
< ±0.2% of range ±2% of setting for ±100 pA range
< ±1% of range ±2% of setting for ±10 pA range
< ±10% of range ±2% of setting for ±1 pA range
Impedance specifications (EIS option only)
Accuracy
10.2.3 1A/48V booster
Control amplifier
Compliance voltage
± 49V
Compliance current
± 1 A
Control voltage
± 48V
Bandwidth (-3dB)
>2MHz