Troubleshooting and Error Messages
78 Bio-Plex Pro and Bio-Plex Pro II Wash Stations
LLD Errors
Examples:
LLD ERROR L
LLD ERROR W
LLD has been inadvertently activated for one channel in your system. This is an
advanced function for which your system is not equipped. All channels should be
deactivated for LLD. The following directions describe how to check and clear all
channels of an LLD setting.
Remember to exit to the Primary Menu before performing this function. (Press EXIT
until the word EXIT disappears.) If this does not solve the problem, please contact
technical support.