WARNINGS and Advice
Before you get started for the first time on this cell sorter consult the responsible
people for proper training
For the moment, bacterial pathogens, cell lines or any other agent of Biosafety
level 2 or higher are not allowed, as well as radioactively labelled or flammable
samples
Do not forget to check the fluidics before you start operating the cell sorter
Use only MiliQ water in the water container
While transferring a cap assembly to a new container, avoid touching the
assembly to the outside surfaces of the containers. If it is necessary to lay the cap
assemblies down, set only on pre-sterilized surfaces
It is recommended that resuspended adherent samples are mixed at high speed
to prevent clumping
Samples must be filtered prior to running. This will minimize clogs in the sample
tube and nozzle tip
Do not open the nozzle door
Keep the green containment door closed as often as possible during the sorting
procedure
Never try to access the laser beam paths
Any adjustment to the toolbars should be avoid. Only knowledgeable users must
perform alterations to the current settings. Please, refer to the responsible people for
any constraint
Optimal settings are automatically determined during the quality control
procedure. Adjusting any of these values will affect the sorting performance
Do not adjust any setting of the global preferences menu
Please confirm the bead lot before starting the quality control procedure. If the
bead lot is not in the dropdown list, please refer to the responsible people
Use always beads from the same lot to perform one quality control procedure
When the remaining run time is down to 5 minutes, the system will automatically
shut down. The fluidics will need to be refilled and the start-up procedure must be
performed. To prevent this, perform the fluidics “hot swap”.