Other Guidelines
Swapping Fluidics During Sorting: “Hot Swap”
New containers or bottles should be prepared before clicking the Swap Fluidics
button, as the swap time is limited to only 3 minutes before the system starting
to shut down due to low fluid levels.
Read carefully all procedure before starting the “Hot Swap
1. Click in the Swap Fluidics button of the Setup and Maintenance tab. The
Swap Fluidics window will open