9. Empty the waste and refill the fluidics containers using the “Hot Swap”
procedure (See Swapping Fluidics During Sorting: “Hot Swap”)
Final Set Up: Shut Down
The shutdown procedure is only recommended for the end of day’s use. If you are
not the last user, please perform a low-pressure cleaning procedure. See Final Set Up:
Low-Pressure Cleaning to do so.
1. Fill a 5 ml tube with a minimum of 3 ml of cleaning solution (70% ethanol)
2. Load the tube into the sample station and move it to the run position
3. In the Set Up and Maintenance tab, click in the Clean System button
4. A Clean Warning dialog will open with a prompt warning for the fact that the
tube solution will be rapidly consumed. Click in the Yes option
5. In the Clean System Modes window, select the High Pressure Clean option and
click in the checkmark button
6. A Clean Warning dialog will open with a prompt warning for the fact that the
high-pressure cleaning will leave the sorter uncalibrated. Click in the Yes option
7. A Clean System process window will be displayed