EasyManua.ls Logo

BIO RAD S3e - Available Techniques; Set Up; Sample Preparation

BIO RAD S3e
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
Operate quickly when performing the “hot swap” procedure, as time is limited
to three minutes for the operation. Bring containers of water or sheath fluid in advance.
The maximum total volumes for a 5 and 1.5 ml collection tubes are 4 and 1 ml,
respectively. Going above these will cause spilling and risk contamination
You must empty the waste and refill the fluidic containers at the end of the use
Perform a high-pressure cleaning protocols if you are the last user of the day.
Otherwise, perform a low-pressure cleaning procedure
Use only 70% ethanol and MiliQ water as cleaning solutions
Please let us know if some solution or fluidic is running out
Make sure to save all the data to your own storage device or cloud as we
routinely wipe out the computer’s data
Do not forget to register your utilization in the logbook. Remember that future
maintenance costs and repairs will be supported by the groups according to the
logbook data
This equipment is only available for booking in the ITQB intranet
Available Techniques
Fluorescence-Activated Cell Sorting (FACS)
Flow Cytometry Analysis
Set Up
Sample Preparation
Cells Suspension
To avoid clogs when working with cells, resuspend cells into a single cell suspension
and always use a 40 µm or smaller filter prior to cell sorting

Related product manuals