Bio-Well Company
www.bio-well.com
6. During menstruation and ± 2 days from it EPI/GDV data can be significantly
perturbed for women.
7. Client does not have long nails (long nails will not allow person to position
fingers correctly on the glass electrode).
*BW 3.0 has an extra space at the top of the glass electrode and more flat
orientation of the glass surface that allows to capture people even with long nails.
8. Ask about any medication that client takes as it greatly affects the results.
NB! If your client has washed his/her hands (or had to wash them) before the
examination please ask him/her to dry them properly and wait for 15-20 minutes to
restore the biochemistry of the skin.
Rules of capturing fingers
Try your best to make the inscribed ellipse of each finger as close to the circle as
it is possible and position it in the center or upper part of the electrode (this is your
guarantee of correct finger positioning and further on – of more accurate analysis).
Try your best to make capture immediately after the client has put the finger on
the electrode (fingers become sweaty very fast without ventilation and the results will
be distorted).
Bio-Well 3.0 device has an in-built ventilator that is able to ventilate the area
around the finger. If the fingers are hot or sweaty, or the humidity is high in the place
where you work with your Bio-Well 3.0 device – turn the ventilator ON. You don’t need
to turn it OFF – it will be turned off automatically after 1 minute. Ventilator of BW 3.0
device can be turned ON before the start of any Scan – use “Fan ON” button in the top
right corner.
BW 2.0 also has an in-built ventilator, but it ventilates only the internal part of the
device in order to protect it from the excessive humidity in the environment and evade
condensation of water at the internal side of the glass electrode.
NB! Take into consideration that Bio-Well GDV Camera 1.0 is going to the stand-by
mode after 10 seconds since last capture. It will need about 5 seconds to react on your
click on the “Capture” button in order to return to the active mode. BW 2.0 & 3.0 do not
go to a stand-by mode and are always ready to capture.
If the fingers are very hot or sweaty, please clean (for example with napkin) each
finger right before capture and the electrode after each finger.