Rev. 1.4 10. SPECIFICATION
200
BM1 User’s Manual
· Battery
- Li-ion battery
- Battery status display
- Operating time:4.5hours(with fully charged Battery)
Graphical and Tabular Trends
· Table Trend
- Memory Storage:128 hours
- Data Interval:15 sec
- Display Interval:1MIN, 5, 15, 30, 1HR
· Graphical Trend
- Display Period:30MINS, 60, 90, 3HRS, 6, 12
SpO
2
capacity
· Saturation Range: 0% to 100% oxygen proportion
· Pulse Rate Range: 30 to 254 bpm
· SpO2 accuracy: 70% to 100% ±2 digits, 0% to 69% unspecified
· pulse accuracy: ±2 bpm
· Sensor Red 660nm, 2mW (typical)
Infrared 880nm, 2-2.4mW (typical)
· Minimum Signal: 0.05% modulation (Low perfusion level performance and
Amplitude limitation validation using FLUKE Index 2 Oximetry Simulator)
NIBP capacity
· Technique: Oscillometric
· Measurement mode:
- Manual: Single Measurement
- Auto: automatic Intervals of 1MIN, 2, 3, 4, 5, 10, 15, 20, 30, 1HR, 2, 4, 8
· Pressure Range: 0 to 300 mmHg(Accuracy ±3mmHg)
· Pulse Rate : 30 to 220BPM
· Blood Pressure Measurement Range:
- Adult Pressure : 20 to 260 mmHg
- Pediatric Pressure : 20 to 160 mmHg
- Neonate Pressure : 20 to 130 mmHg
· Accuracy : Meets accuracy requirements of ANSI/AAMI SP10:1992 AND 2002