Rev. 1.4 1.BASIC
BM1 is classified as follows:
- BM1 classifies as Class I, BF & CF concerning electric shock. It is not proper to operate this
Equipment around combustible anesthetic or dissolvent.
- Safety level is class I regarding IEC/EN 60601-1 and the Limit of Nose is Class B concerning
IEC/EN60601-1-2.
Equipment Connection
For measurements in or near the heart we recommend connecting the monitor to the potential
equalization system. Use the green and yellow potential equalization cable and connect it to the pin
labeled with the symbol .
In the hospital, doctors and patients are exposed to dangerous, uncontrollable compensating
currents. These currents are due to the potential differences between connected equipment The
safety solution to the problem is accomplished with EN60601-1;1993.
Biocompatibility
When used as intended, the parts of the product described in this operator manual, including
accessories that come in contact with the patient during the intended use, fulfill the biocompatibility
requirements of the applicable standards. If you have questions about this matter, please contact
BIONET or its representatives.
Maintenance and Washing Equipment Connection
There are various methods to clean the BM1 and its accessories. Please follow the methods
mentioned below to avoid unnecessary damage or contamination to the Equipment.
We do not repair free of charge regardless of warranty period if it is contaminated or damaged with
using dangerous material not approved for washing.