EasyManua.ls Logo

BIOTRONIK EVIA - Page 31

BIOTRONIK EVIA
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30 After Implantation
Rate decrease The decrease of basic rate and magnet rate is defined as follows:
In the following pacing modes the pacing rate decreases by 11%: DDD(R);
DDT(R); DOO(R) VDD(R); VDI(R); VVI(R), VVT(R) AAI(R); AAT(R); AOO(R).
In the pacing modes DDI(R) and DVI(R), only the VA interval is extended by 11%.
This reduces the pacing rate by 4.5 to 11%, depending on the configured AV
delay.
Magnet response at ERI After reaching ERI pacing is performed as follows after applying the magnet or pro-
gramming head:
Expected service time
after ERI
The information is based on a lead impedance of 500 Ohm at 100% pacing and
the data of the battery manufacturer.
For a lead impedance of 300 Ohm instead of 500 Ohm, these times decrease by
max. 30%.
Parameter with high pulse energy: 110 ppm; 4.6 V; 1.5 ms; 500 Ohm
Parameter with low pulse energy: 30 ppm; 0.2 V; 0.1 ms; 500 Ohm
Dual-chamber implant in DDDR mode; single-chamber implant in
AAIR/VVR mode:
[in months]
Magnet mode
Cycles 1 to 10: After 10th cycle:
Automatically Asynchronous with rate at
80 ppm
Synchronous with basic rate
reduced by 4.5 to 11%
Asynchronous Asynchronous with rate at
80 ppm
Asynchronous with rate at
80 ppm
Synchronous Synchronous with basic rate
reduced by 4.5 to 11%
Synchronous with basic rate
reduced by 4.5 to 11%
ERI to EOS interval Standard program With high pulse
energy
With low pulse
energy
Mean value 8 8 8
Minimum value 6 6 6

Related product manuals