44 Parameters
Preset Programs SR(-T)
Standard and safe program Only the auto-initialization function is activated as a factory setting. All the other
functions of the standard program are deactivated.
Parameter Standard program Safe program
Mode (after auto initialization: VVI) VVI VVI
In the AAI mode, the safe program is also
AAI.
Basic rate 60 ppm 70 ppm
Night program OFF OFF
Rate hysteresis OFF OFF
Magnet response AUTO AUTO
Pulse amplitude 3.0 V 4.8 V
Pulse width 0.4 ms 1.0 ms
Sensitivity AUTO 2.5 ms
Refractory period 250 ms 300 ms
Lead configuration, automatically determined and set
Pacing polarity Unipolar Unipolar
Sensing polarity Unipolar Unipolar
Automatic lead check ON ON
Active capture control ATM OFF
IEGM recording ON OFF
Home Monitoring OFF OFF