42 Parameters
Preset Programs DR(-T)
Standard and safe program Only the auto-initialization function is activated as a factory setting. All the other
functions of the standard program are deactivated.
Parameter Standard program Safe program
Mode (after auto initialization:
DDD-CLS)
DDDR VVI
Basic rate 60 ppm 70 ppm
Night program OFF OFF
Rate hysteresis OFF OFF
Upper rate 130 ppm
Dynamic AV delay Low
AV hysteresis OFF
Sense compensation –45 ms
AV safety delay 100 ms
Far-field protection after V
s
100 ms
Far-field protection after V
p
150 ms
Ventricular blanking period after A
p
32 ms
PMT protection ON
VA criterion 380 ms
Magnet response AUTO AUTO
Pulse amplitude A 3.0 V
Pulse amplitude V 3.0 V 4.8 V
Pulse width A 0.4 ms
Pulse width V 0.4 ms 1.0 ms
Sensitivity A AUTO
Sensitivity V AUTO 2.5 mV
Refractory period A AUTO
Refractory period V 250 ms 300 ms
Mode switching ON
Onset criterion 5-out-of 8
Resolution criterion 5-out-of 8
Intervention rate 160 ppm
Switches to DDIR
Basic rate with mode switching +10 ppm
Rate stabilization with
mode switching
OFF
PVARP AUTO AUTO
PVARP after PVC 400 ms