EasyManua.ls Logo

BIOTRONIK EVIA - Replacement indications

BIOTRONIK EVIA
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29 After Implantation
Replacement Indications
Pacemaker operational
status indications
The time span from the beginning of service (BOS) to elective replacement indica-
tion (ERI) is determined by, among others, the following:
• Battery capacity
• Lead impedance
• Pacing program
• Pacing to inhibition ratio
• Pacemaker circuit properties
The following are the defined pacemaker operational statuses:
ERI activation ERI detection is automatically activated after the following events:
• Successful auto-initialization
• Storage for longer than 24 months
ERI display ERI is displayed as follows:
• By a defined decrease in the basic rate as well as the magnet rate
• After interrogation of the pacemaker
Change of the pacing mode
with ERI
From dual-chamber modes, the pacemaker switches to single-chamber pacing.
This replacement mode depends on the programmed mode and is displayed on the
programmer.
Deactivated functions
with ERI
The following functions are deactivated:
• Atrial pacing
• Night program
• Rate adaptation
• Atrial and ventricular active capture control
• Rate fading
• Atrial overdrive pacing
• IEGM recordings
• Statistics
• Home Monitoring
• Rate hysteresis
• Ventricular pacing suppression
BOS Beginning of Service Battery is in good condition; normal
follow-up.
ERI Elective Replacement
Indication
The replacement time has been reached.
The pacemaker must be replaced.
EOS End of Service End of service time with regular pace-
maker activity.