38 Parameters
Pacing and Sensing DR(-T)
Pulse amplitude
and pulse width
Sensitivity
Atrial capture control
Ventricular capture control
Parameter Range of values Standard
Pulse amplitude A 0.2 ... (0.1) ... 6.2; 7.5 V 3.0
Pulse width A 0.4; 0.5; 0.7; 1.0; 1.2; 1.5 ms 0.4 ms
Pulse amplitude V 0.2 ... (0.1) ... 6.2; 7.5 V 3.0 V
Pulse width V 0.4; 0.5; 0.7; 1.0; 1.2; 1.5 ms 0.4 ms
Parameter Range of values Standard
Sensitivity A AUTO
0.1 ... (0.1) ... 1.5 ... (0.5) ... 7.5 mV
AUTO
Sensitivity V AUTO
0.5 ... (0.5) ... 7.5 mV
AUTO
Parameter Range of values Standard
Atrial capture control ATM (monitoring only); OFF OFF
Min. amplitude 0.5 ... (1) ... 4.8 V 1.0 V
Threshold test start 2.4; 3.0; 3.6; 4.2; 4.8 V 3.0 V
Safety margin 0.5 ... (1) ... 1.2 V 1.0 V
Search time Interval; time of day Interval
Interval 0.1; 0.3; 1; 3; 6; 12; 24 h 24 h
Time of day 00:00 ... (15 min) ... 23:45 02:00
Parameter Range of values Standard
Ventricular capture control ON; ATM (monitoring only); OFF ON
Min. amplitude 0.7 V 0.7 V
Threshold test start 2.4; 3.0; 3.6; 4.2; 4.8 V 3.0 V
Safety margin 0.3 ... (1) ... 1.2 V 0.5 V
Search time Interval; time of day Interval
Interval 0.1; 0.3; 1; 3; 6; 12; 24 h 24 h
Time of day 00:00 ... (15 min) ... 23:45 hh:mm 02:00 hh:mm