49 Technical Data
Battery Data
Battery type characteristics The following data is entered by the manufacturer:
Power consumption The implant has the following power consumption:
Average service time Average service times are precalculated using the battery manufacturer's technical
specifications, a basic rate of 60 ppm and the setting of different pulse amplitudes
and lead impedances.
Service times DR(-T) For dual-chamber implants, the following times (in years) result:
Manufacturer GREATBATCH, INC.
Clarence, NY 14031, USA
LITRONIK GmbH
01796 Pirna, Germany
Battery type GB 8431 GB 2596 LiS 3150 LiS 3150M
System LiJ Ag/SVO/CFx
QMR
®
LiJ LiMn0
2
Implant DR
SR
DR-T
SR-T
DR
SR
DR-T
SR-T
Battery voltage
at BOS
2.8 V 3.0 V 2.8 V 3.1 V
Open-circuit volt-
age
2.8 V 3.0 V 2.8 V 3.1 V
Nominal capacity 1.3 Ah 1.3 Ah 1.3 Ah 1.2 Ah
Usable capacity
until EOS
1.2 Ah 1.1 Ah 1.2 Ah 1.0 Ah
Power consumption DR(-T) SR(-T)
BOS, inhibited 6 μA 6 μA
BOS, 100% pacing 13 μA 9 μA
Amplitude Impedance
[Ohms]
Pacing
10% 50% 100%
DR-T DR DR-T DR DR-T DR
1.5 V 500 19.3 17.8 17.2 15.8 15.0 13.9
1000 19.7 18.0 18.3 16.9 17.0 15.7
2.5 V 500 17.8 16.6 13.6 12.8 9.4 10.0
1000 18.8 17.4 15.3 14.9 12.5 12.7
3.0 V 500 17.3 15.8 11.8 10.9 8.5 7.8
1000 18.5 16.9 14.6 13.4 11.6 10.7
3.5 V 500 16.5 15.1 10.4 9.7 7.2 6.6
1000 18.0 16.5 13.4 12.4 10.2 9.4
5.0 V 500 12.4 12.1 5.4 6.2 3.2 3.8
1000 15.2 14.6 8.3 9.2 5.3 6.3