50 Technical Data
Service times SR(-T) For single-chamber implants, the following times (in years) result:
Shortening
of the service time
after long storage period
Depending on the storage period, the service time from the beginning of service
BOS to the replacement time ERI decreases as follows:
• After 1 year:
— DR(-T) by 6 months
— SR(-T) by 8 months
• After 1.5 years:
— DR(-T) by 9 months
— SR(-T) by 12 months
Amplitude Impedance
[Ohms]
Pacing
10% 50% 100%
SR-T SR SR-T SR SR-T SR
1.5 V 500 23.3 21.3 21.7 19.8 19.9 18.3
1000 23.6 21.5 22.7 20.7 21.6 19.8
2.5 V 500 22.1 20.3 17.8 17.3 14.3 14.6
1000 22.8 21.0 20.2 19.2 17.5 17.2
3.0 V 500 21.7 19.8 16.8 15.5 13.2 12.2
1000 22.7 20.7 19.5 17.8 16.6 15.3
3.5 V 500 20.9 19.1 15.3 14.2 11.5 10.7
1000 22.3 20.3 18.4 16.9 15.1 13.9
5.0 V 500 17.2 16.3 9.1 10.1 5.7 6.8
1000 19.8 18.7 12.8 13.7 8.9 10.2