EasyManua.ls Logo

BIOTRONIK EVIA - Tolerances of Parameter Values

BIOTRONIK EVIA
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
45 Parameters
Tolerances of Parameter Values
DR(-T)
Parameter Range of values Tolerance
Basic rate 30 ... 100 ppm +/-1.5 ppm
102 ... 195 ppm +/-2.0 ppm
200 ppm +0.0/-3.0 ppm
Basic interval 1000 ms +/-20 ms
Magnet rate 90 ppm +/-1.5 ppm
Magnet interval 664 ms +/-20 ms
AV delay 15 ... 350 ms +20/-5 ms
A/V pulse amplitude 0.2 V +/-0.10 V
0.3 ... 7.5 V +20/-25%
A/V pulse duration 0.1 ... 0.4 ms +/-0.04 ms
0.5 ... 1.0 ms +/-0.10 ms
1.25 ... 1.5 ms +/-0.15 ms
Sensitivity A
45502-2-1 Delta pulse
0.1 ... 0.5 mV +/-0.10 mV
0,6 ... 7.5 mV +/-20%
Sensitivity V
45502-2-1 Delta pulse
0.5 ... 7,5 mV +/-20%
Refractory period A 300 ... 775 ms +10/-30 ms
Refractory period V 200 ... 500 ms +10/-30 ms
PVARP 175 ... 600 ms +10/-30 ms
PVARP after PVC 325 ... 600 ms +10/-30 ms
Max. activity rate 80 ... 100 ppm +/-1.5 ppm
105 ... 160 ppm +/-2.0 ppm
Upper rate 90 ... 190 ppm +/-2.0 ppm
200 ppm +0/-2.0 ppm
High rate protection 200 ppm +20/-0 ppm
Lead impedance 100 ... 200 Ohm +/-50 Ohm
201 ... 2500 Ohm +/-25%