41 Parameters
Rate Adaptation
Closed Loop Stimulation
rate adaptation
CLS modes:
Rate adaptation
via accelerometer
R modes:
Parameter Value range Standard
Max. CLS rate 80 ... (5) ... 160 ppm 120 ppm
CLS response Very low; low; medium; high;
very high
Medium
Resting rate control OFF;
+10 ... (10) ... +50 ppm
+20 ppm
CLS required Yes; no No
Parameter Range of values Standard
Sensor gain 1 ... 23 4
Max. activity rate 80 ... (5) ... 160 ppm 120 ppm
Automatic gain OFF; ON ON
Sensor threshold Very low; low; medium; high; very
high
Medium
Rate increase 1; 2; 4; 8 ppm/cycle 4 ppm/cycle
Rate decrease 0.1; 0.2; 0.5; 1.0 ppm/cycle 0.5 ppm/
cycle
Rate fading OFF; ON OFF