34 Parameters
Timing DR(-T)
Basic rate day/night
Rate hysteresis
AV delay
AV hystereses
Parameter Range of values Standard
Basic rate 30 ... (1) ... 90 ... (2) ... 122 ... (3) ... 140
... (5) ... 200 ppm
60 ppm
Night rate OFF;
30 ... (1) ... 90 ... (2) ... 122 ... (3) ... 140
... (5) ... 200 ppm
OFF
Night begins 00:00 ... (10 min) ... 23:59 hh:mm 22:00 hh:mm
Night ends 00:00 ... (10 min) ... 23:59 hh:mm 06:00 hh:mm
Parameter Range of values Standard
Rate hysteresis OFF;
-5 ... (-5) ... -90 ppm
OFF
Repetitive hysteresis OFF; 1 ... (1) ... 15 OFF
Scan hysteresis OFF; 1 ... (1) ... 15 OFF
Parameter Range of values Standard
AV delay Low; medium; high; fixed;
individual
Low
AV delay 15 ... (5) ... 350 ms
(in 6 rate ranges)
180 ms
Sense compensation OFF;
-10 ... (5) ... -120 ms
-45 ms
AV safety delay 100 ms 100 ms
Parameter Range of values Standard
AV hysteresis mode OFF;
Negative, low; medium; high;
IRSplus
OFF
Positive repetitive
AV hysteresis
OFF;
1 ... (1) ... 10
OFF
Negative repetitive
AV hysteresis
OFF;
1 ... (1) ... 15 ... (5) ... 100 ... (10)
... 180
OFF
AV scan hysteresis OFF;
1 ... (1) ... 10
OFF