37 Parameters
Timing SR(-T)
Basic rate day/night
Rate hysteresis
Upper rate
Refractory period
Parameter Range of values Standard
Basic rate 30 ... (1) ... 90 ... (2) ... 122 ... (3) ... 140
... (5) ... 200 ppm
60 ppm
Night rate OFF
30 ... (1) ... 90 ... (2) ... 122 ... (3) ... 140
... (5) ... 200 ppm
OFF
Night begins 00:00 ... (10 min) ... 23:59 hh:mm 22:00 hh:mm
Night ends 00:00 ... (10 min) ... 23:59 hh:mm 06:00 hh:mm
Parameter Range of values Standard
Rate hysteresis OFF
-5 ... (-5) ... -90 ppm
OFF
Repetitive hysteresis OFF; 1 ... (1) ... 15 OFF
Scan hysteresis OFF; 1 ... (1) ... 15 OFF
Parameter Range of values Standard
Upper rate 90 ... (10) ... 200 ppm 130 ppm
Parameter Range of values Standard
Refractory period 200 ... (25) ... 500 ms 250 ms