35 Parameters
Ventricular
pacing suppression
Upper rate
Mode switching
Refractory periods
Parameter Range of values Standard
V
p
suppression OFF; ON OFF
Pacing suppression after
consecutive V
s
1 ... (1) ... 8 6
Pacing supports after
X-out-of-8 cycles
1; 2; 3; 4 3
Parameter Range of values Standard
Upper rate 90 ... (10) ... 200 ppm 130 ppm
Atrial upper rate OFF;
240 ppm
240 ppm
Parameter Range of values Standard
Mode switching OFF; ON ON
Intervention rate 100 ... (10) ... 250 ppm 160 ppm
Switch to (mode) DDI;
DDI(R) when permanent DDD(R)
VDI;
VDI(R) when permanent VDD(R)
DDIR
VDIR
Onset criterion 3 ... (1) ... 8 5
Resolution criterion 3 ... (1) ... 8 5
Change of the basic rate
with mode switching
OFF;
+5 ... (5) ... +30 ppm
+10 ppm
Rate stabilization with
mode switching
OFF; ON OFF
Parameter Range of values Standard
Atrial refractory period AUTO AUTO
Atrial refractory period in
the modes AAI(R); AAT(R);
DDT
300 ... (25) ... 775 ms 350 ms
PVARP AUTO;
175 ... (5) ... 600 ms
AUTO
PVARP after PVC PVARP + 150 ms (max: 600 ms)
is automatically programmed
400 ms
Ventricular
refractory period
200 ... (25) ... 500 ms 250 ms