39 Parameters
Lead configuration
IEGM recordings
Rates for statistics
Parameter Range of values Standard
Pacing polarity A Unipolar; bipolar Unipolar
Pacing polarity V Unipolar; bipolar Unipolar
Sensing polarity A Unipolar; bipolar Unipolar
Sensing polarity V Unipolar; bipolar Unipolar
Parameter Range of values
IEGM recordings 20 (quantity); each max. 10 s
Types of IEGM recordings High atrial rate (HAR)
Mode switching (MSW)
High ventricular rate (HVR)
Patient triggered (Pt.)
IEGM recording prior to event 0; 25; 50; 75; 100%
IEGM signal Filtered; unfiltered
Parameter Range of values Standard
High atrial rate (HAR) 100 ... (5) ... 250 bpm
600; 572 ... 245; 240 ms
200 bpm
300 ms
High ventricular rate (HVR) 150 ... (5) ... 200 bpm
400; 378 ... 308; 300 ms
180 bpm
333 ms
HVR counter 4; 8; 12; 16 8