BT-770 Operation manual
114
P/N : 770-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
Source
Alarm message
P
Causes and countermeasures
monitor's specifications, and check whether
there are special reasons affecting ambient
pressure.
NIBP over range L
The measured blood pressure of the patient
exceeds the measuring range.
NIBP signal weak L
Patient’s pulse may be weak or cuff is too loose.
Check the condition of the patient, and place the
cuff in a suitable position. If the failure persists,
replace the cuff.
NIBP signal unstable
L
Excessive movement may result in too much
motion artifact or interference in the signal
during measurement.
NIBP signal saturated
L
Motion signal amplitude is too large due to
movement and other reasons.
NIBP over pressure
L
Cuff overpressure, and gas blockage may occur;
check the gas path, and then re-measure.
Module reset failed L
NIBP module reset error; check the gas path is
blocked, and then restart the measurement.