P/N : 770-ENG-OPM-EUR-D03
Causes and countermeasures
monitor's specifications, and check whether
there are special reasons affecting ambient
pressure.
NIBP over range L
The measured blood pressure of the patient
exceeds the measuring range.
NIBP signal weak L
Patient’s pulse may be weak or cuff is too loose.
Check the condition of the patient, and place the
cuff in a suitable position. If the failure persists,
replace the cuff.
Excessive movement may result in too much
motion artifact or interference in the signal
during measurement.
Motion signal amplitude is too large due to
movement and other reasons.
Cuff overpressure, and gas blockage may occur;
check the gas path, and then re-measure.
Module reset failed L
NIBP module reset error; check the gas path is
blocked, and then restart the measurement.