EasyManua.ls Logo

BK Precision 9200B Series - MEASUREMENT Commands

BK Precision 9200B Series
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
43
Return parameter: <NRf>
Return parameter unit: V
MEASure[:SCALar]:CURRent[:DC]?
This command is used to read the actual current.
Command syntax: MEASure[:SCALar]:CURRent[:DC]?
Parameter: None
Return parameter: <NRf>
Return parameter unit: A
Example: MEAS:CURR?
FETCh:CURRent?
This command is used to read the current which is in the sample cache. After sending the
command, the readings will be sent to the computer. This command does not affect the
instrument settings. This command does not trigger a measurement operation, and queries
only the latest available reading. Before reading the new reading, the command returns the old
readings.
Command syntax: FETCh:CURRent?
Return parameter: <NRf>
MEASure[:SCALar]:POWer[:DC]?
This command is used to query the actual output power.
Command syntax: MEASure[:SCALar]:POWer[:DC]?
Parameter: None
Return parameter: <NRf>
Return parameter unit: W
Example: MEAS:POWer?
FETCh:POWer?
This command is used to read the power which is in the sample cache. After sending the
command, the readings will be sent to the computer. This command does not affect the
instrument settings. This command does not trigger a measurement operation, and queries
only the latest available reading. Before reading the new reading, the command returns the old
readings.
Command syntax: FETCh:POWer?
Return parameter: <NRf>
MEASure:DVM?
This command is used to read the latest measured value of the DVM meter.
Command syntax: MEASure[:SCALar]:DVM[:DC]?
Return parameter: <NRf>
FETCh:DVM?
This command is used to read the latest data of DVM meter from the sampling buffer.
www.calcert.com sales@calcert.com1.888.610.7664
0
5
10
15
20
25
30

Table of Contents