YH-600B Pro User Manual V3.0
20 / 88
WARNINGS!
Change the measured finger regularly according to the
patient's conditions after prolonged use. It must be moved
every 4 hours or more often and reapplied to another site
according to the users skin integrity and circulatory status.
Excessive ambient light, excessive motion, use of
intravascular dyes, poorly perfused finger, extreme finger
sizes or improper placement may affect the accuracy of the
measurement.
Nail polish or false nails may cause erroneous measurement
results.
Overly low blood pressure, overly low systolic blood pressure,
severe anemia, or hypothermia may cause erroneous
measurement results.
The pulse oximeter sensor is designed for use with this
device only.
Check the compatibility of the device and pulse oximeter
sensor before use; otherwise it may cause injury to the
patient.
A functional tester cannot be used to assess the accuracy of
the SpO
2
measurement result.
Do not use the pulse oximeter sensor during MRI scanning.
Do not use a broken pulse oximeter sensor.
Do not immerse the pulse oximeter sensor in liquid, as it
causes short circuit.
The pulse oximeter sensor may not work on cold extremities
due to reduced circulation. Warm or rub the finger to increase
circulation, or reposition the sensor.
If the pulse oximeter sensor is shared by multi-users, the
users may be exposed to the risk of cross-infection.
Blood flow must not be impeded during the recording process.
The pulse oximeter sensor should be applied snug and
comfortable, but not too tight or subject to any excessive
pressure.