12. Cleaning and Disinfection
12.1 Cleaning
YH-600B Pro was designed to be simple and easy-to-use. Your
YH-600B Pro will operate reliably with good care and regular
maintenance.
Cleaning instruction:
Clean YH-600B Pro Main Device
① Turn off YH-600B Pro and take battery out of YH-600B Pro
main device.
② Scrub YH-600B Pro main device gently with a damp gauze
pad.
③ Dry YH-600B Pro main device with a dry soft gauze pad.
Clean the Respiration Kit
① Detach the Respiration Kit and the CPAP pressure cannula.
② Wipe the Respiration Kit gently with a damp gauze pad.
③ Dry the Respiration Kit with a dry soft gauze pad.
Clean the USB Data Cable
① Disconnect the USB data cable from YH-600B Pro main
device.
② Scrub the USB data cable gently with a damp gauze pad.
③ Dry the USB data cable with a dry soft gauze pad.
Clean the Pulse Oximeter Sensor
The pulse oximeter sensor may be surface-cleaned by wiping it