YH-600B Pro User Manual V3.0
34 / 88
with a solution, such as 70% isopropyl alcohol.
① Detach the pulse oximeter sensor from YH-600B Pro main
device.
② Saturate a clean, dry gauze pad with the cleaning solution.
Wipe all surfaces of the pulse oximeter sensor with this gauze
pad.
③ Saturate another clean, dry gauze pad with potable water.
Wipe all surfaces of the pulse oximeter sensor with this gauze
pad.
④ Dry the pulse oximeter sensor by wiping all surfaces with a
clean, dry gauze pad.
Clean the CPAP Pressure Tube
① Detach the CPAP pressure tube from the respiration kit.
② Immerse the CPAP pressure tube completely in a solution,
such as 70% isopropyl alcohol.
③ Rinse the CPAP pressure tube with potable water.
④ Dry the CPAP pressure tube by wiping all surfaces with a clean,
dry gauze pad and air dry the lumen of the CPAP pressure tube.
Cleaning the Wrist Belt
① Wash the wrist belt before first time use, as the dye may run.
② Detach the wrist belt from the main device.
③ It is important to always hand-wash the wrist belt.
④ Wash the wrist belt in warm (approx. 40°C) and soapy water.
Use pure soap.
⑤ Rinse the wrist belt well with potable water and allow it to air
dry out of direct sunlight.