Disinfect the Wrist Belt
① Detach the wrist belt from the main device.
② Immerse the wrist belt completely in CaviCide
®
solution for a
minimum of 5 minutes at 20°C (68˚F) or higher to destroy all
pathogenic microorganisms.
③ Rinse the wrist belt with potable water. Keep the wrist belt
totally immersed for a minimum of 1 minute in duration.
④ Remove the wrist belt and discard the rinse water. Always use
fresh volumes of water for each rinse. Do not reuse the water for
rinsing or any other purpose. Repeat rinsing the wrist belt for two
additional times, for a total of three wipes to remove CaviCide
®
solution residues.
⑤ Air dry the wrist belt out of direct sunlight.
Sterilization of the device and its components other than
recommended is not permitted.
WARNING!
• The device shall not be serviced or maintained while in use
with a patient.
13. Disposal
Dispose of the device and its package in accordance with local
regulations.
Do not dispose device as household waste. To ensure proper
disposal of the device, consult an authorized and certified
electronic scrap recovery firm. You can find out their address from