YH-600B Pro User Manual V3.0
36 / 88
Saturate another clean, dry gauze pad with potable water.
Wipe all surfaces of the pulse oximeter sensor with this gauze
pad.
Discard the used potable water. Always use fresh volumes of
water for each wipe. Do not reuse the water for wipe or any other
purpose. Repeat rinsing the pulse oximeter sensor for two
additional times, for a total of three wipes to remove CaviCide
®
solution residues.
Dry the pulse oximeter sensor by wiping all surfaces with a
clean, dry gauze pad.
Disinfect the CPAP Pressure Tube
Detach the CPAP pressure tube from the respiration kit.
Immerse the CPAP pressure tube completely, filling all lumens
and eliminating air pockets, in CaviCide
®
solution for a minimum
of 5 minutes at 20°C (68˚F) or higher to destroy all pathogenic
microorganisms.
Rinse the CPAP pressure tube with potable water. Keep the
CPAP pressure tube totally immersed for a minimum of 1 minute.
Manually flush all lumens with large volumes (not less than 100 ml)
of rinse water.
Remove the CPAP pressure tube and discard the rinse water.
Always use fresh volumes of water for each rinse. Do not reuse
the water for rinsing or any other purpose. Repeat rinsing the
CPAP pressure tube for two additional times, for a total of three
wipes to remove CaviCide
®
solution residues.
Dry the CPAP pressure tube by wiping all surfaces with a clean,
dry gauze pad and air dry the lumen of the CPAP pressure tube.