2 Safety
900-301_IFU-V2.2_11316-S0-20131023-EN Operating Manual ARC 250 / 303 17
2.2. General safety instructions
Ensure that no electronic devices that are subject to interference
from electromagnetic fields are set up in the vicinity of the HF
device.
Observe the instructions on electromagnetic compatibility provided
in section EMC, page 86.
Always connect the HF device to a mains power system with a
protective earth lead in order to prevent electric shock.
Additional devices that are connected to electrical medical devices must
satisfy relevant IEC or ISO standards (e.g. IEC 60950 for data
processing devices). Furthermore, all configurations must comply with
the standardised requirements for medical systems (see IEC 60601-1-1
or Section 16 of the 3rd edition of IEC 60601-1 as relevant). Anyone
who connects additional devices to electrical medical devices is
automatically a system configurator and thus responsible for meeting
standardised system requirements. Please note that local laws prevail
over the aforementioned standard requirements. In case of questions,
please contact your local dealer or Techical Service, see
section Technical service, page 59.
To protect personnel, BOWA recommends the use of a smoke evacuator
to extract electrosurgical smoke, e.g. BOWA SHE SHA.